Service & Repair
Fast diagnosis and expert fixes, 24 hours a day, on virtually any make or model.
Refrigerated centrifuges · Repair & service
A refrigerated centrifuge is a refrigeration system wrapped around a spinning rotor. When the chamber will not hold 4°C, samples warm during the run and the results follow. We service the refrigeration side of refrigerated centrifuges: compressor, condenser, chamber temperature control, and the sensors that report it.
Centrifuge not holding temperature? Call and dispatch triages it with you now. Or send an emergency request and a dispatcher calls you back.
The refrigeration subsystem of benchtop, floor-model, and high-speed refrigerated centrifuges: compressor and condenser, chamber evaporator, refrigerant charge, chamber temperature sensor and control, and the condenser airflow path. Rotor, drive, and imbalance systems are the instrument service provider's domain; we coordinate with them when a visit touches both.
Centrifuges live under benches where condensers pack with dust and airflow is poor, so head pressure climbs and the chamber cannot pull down. Long high-speed runs generate heat the refrigeration has to remove; a weak compressor shows up as a chamber that reaches 4°C empty but not under load. A drifting chamber sensor reports a temperature the samples never saw.
Condenser airflow and head pressure are checked first, then compressor performance under a loaded run, then the chamber sensor against a reference.
Compressors, condensers and fans, chamber evaporators, refrigerant charge, and chamber temperature sensors and control boards on the refrigeration side.
Chamber temperature verified against a traceable reference at rest and during a run, and the condenser path cleaned and checked.
A written record of what failed and what was done, and a plain repair-or-replace recommendation when a unit is past its useful life.
Rotor, drive, and safety systems stay with your instrument service provider, and we coordinate when both are involved. We do not rent centrifuges.
We verify chamber temperature against a traceable reference at rest and during a run, clean and check the condenser path, and document the result for the lab's equipment file.
Brands
Linked brands have their own page. Not on the list? Ask. We service most makes and models.
Where
Ten Southern California counties from our Huntington Beach headquarters, and beyond by arrangement.
Los Angeles, Orange, Riverside, San Bernardino, San Diego, and Ventura counties are our core area. Santa Barbara, San Luis Obispo, Kern, and Imperial counties are on a regular route. Equipment that is already down jumps the queue, and our line is answered 24/7.
Outside Southern California we work by arrangement for contract and long-standing customers, and we say plainly what we can do and how soon.
Also from one team
Fast diagnosis and expert fixes, 24 hours a day, on virtually any make or model.
Scheduled upkeep on a cadence that fits each unit, so wear is found on a visit instead of as downtime.
Refrigerated centrifuges FAQs
The refrigeration system: compressor, condenser, chamber cooling, and temperature control. Rotor, drive, and safety systems stay with your instrument service provider, and we coordinate when both are involved.
Heat from the run is outrunning the refrigeration, usually because the condenser is dirty or the compressor is weak. Both are diagnosable in one visit.
Ready when you are
Request service online in under a minute, or call. Our line is answered 24/7.
24/7 emergency service · Dispatch confirms every request by phone.