A temperature mapping study answers one question: does every location in this chamber stay within the required range under the conditions it will actually operate in? The study itself is straightforward. What makes it go well or badly is what happens before the probes go in. This guide covers the decisions and preparation that let a study pass the first time.
What a mapping study is
Calibration checks a single sensor against a reference. Mapping places many calibrated probes throughout a chamber, records temperature over a defined period, and reports how it varied by location and over time. The result shows hot and cold spots, the effect of door openings and defrost cycles, and whether the unit’s own sensor represents what the product sees.
The output is a report with the raw data, the analysis, and a conclusion against acceptance criteria that were set before the study. That report is what an auditor asks for.
Decide these things first
The most common reason a study has to be repeated is that a decision was made after the data was collected. Settle these before scheduling.
Acceptance criteria
Write down the temperature range every probe must stay within, and for how long an excursion is tolerated, if at all. For a +2°C to +8°C refrigerator the range is usually the label range. For a -80°C freezer it may be a range like -70°C to -90°C. Your product requirements and your SOP set these, not the technician.
Load state
An empty chamber behaves differently from a full one. Air moves freely in an empty unit; a loaded unit has thermal mass and blocked airflow. Decide whether the study is empty, loaded, or both, and if loaded, what a representative load is. Use dummy product of similar mass and packaging if real product cannot be used.
Duration and conditions
How long the probes stay in, whether door openings are simulated, whether a power-loss or door-open challenge is included, and what the room temperature will be. A study run in January in a room that is 30°C in August may not represent worst case.
Twenty-four to 72 hours is a common starting point for a refrigerator or freezer, but your SOPs dictate the duration and conditions, and those depend on what you have contracted us for. Requirements can also come from the manufacturer’s IQ, OQ, and PQ documentation or its mapping recommendations. Many organizations, especially those with European or overseas parents, already have protocols written and approved; it may be a matter of handing them over rather than starting from scratch.
Probe locations
Probes go where temperature is most likely to vary: corners, near the door, top and bottom, close to the evaporator, close to the unit’s own sensor, and in the geometric center. The number of probes depends on the chamber size. The plan should show each location on a diagram so the study can be repeated.
Who approves
Name the person who approves the protocol before the study and the report after it. If that person is on vacation during the study window, move the study.
One point worth being plain about: we do not tell you what your study needs. We draft the protocol and send it to you to approve. Your requirements are yours, and the sign-off is yours. That is how it has to work for the record to mean anything to an auditor.
Prepare the unit
A mapping study records what the unit does. It does not fix anything. If a unit has a known problem, fix it first or the study will document the problem in detail and then need to be repeated.
- Have the unit serviced and its sensors calibrated before mapping if it is due or overdue.
- Clean the condenser and check the gasket. A leaking gasket produces a warm zone near the door that will show up on every probe there.
- Confirm the setpoint is what the protocol says it is, and that alarm limits are set.
- Let the unit run at setpoint long enough to be stable before the study starts. A unit that was moved or restarted the day before is not stable.
- For a walk-in, check the door closer, the strip curtain if there is one, the evaporator fans, and the defrost schedule. Defrost cycles will appear in the data and should be expected, not discovered.
- Do not schedule the study during a period of unusual use: a large delivery, a freezer consolidation, a room renovation.
Prepare the room and the people
- Tell everyone who uses the unit that a study is running, what they may and may not do, and for how long. Taping a notice to the door is not enough; a probe knocked loose by someone reaching for a rack invalidates that location.
- If door openings are part of normal use, decide whether the study includes them at a normal rate or excludes them. Either is defensible; undocumented openings are not.
- Confirm the room’s temperature control will hold during the study and that no HVAC work is scheduled.
- Ideally the unit sits alone in a controlled area for the duration and nobody uses it. That is not always possible; when it is not, the protocol should say how normal use is handled.
- Make sure the probes and logger can be placed without pinching a cable in the door.
Prepare the paperwork
The protocol package we send ahead for approval already includes the unit’s make, model, serial number, asset tag, and location, along with the acceptance criteria, load state, duration, and probe diagram. Your side of the paperwork is:
- Confirm the identification and location in the protocol are correct, and sign the approval before the study date.
- Have recent service and calibration records for the unit on hand.
- Have your SOP for mapping, or the section of your quality manual that governs it, available so the report can reference it.
- File the calibration certificates for the mapping probes, which we supply.
During the study
Leave the unit alone unless the protocol calls for a challenge. Record anything unusual that happens in the room: a power blip, a door left open, a delivery. Those notes explain excursions in the data and save a repeat study.
After the study
The report should give you, for each probe, the minimum, maximum, and mean temperature, the time outside range if any, and the location. It should identify the warmest and coldest locations, compare the unit’s own sensor to the probes, and state clearly whether the acceptance criteria were met. If they were not, the report should say where and by how much, and the next step is a repair or a change in how the unit is used, followed by a new study.
Use the results. If the warmest location is the top shelf near the door, do not store your most sensitive product there. If the unit’s sensor reads colder than the product locations, consider where the monitoring probe sits. The study is only valuable if it changes something.
When to call us
Our temperature mapping and validation services cover refrigerators, freezers, ultra-low freezers, incubators, stability chambers, and walk-in rooms across Southern California, with NIST-traceable probes and reports written for your auditor. If a unit needs service before it can be mapped, the same team can do that first.